Checklist for Set-Up of NIV (Bi-Level NIV)

  • Explain the treatment to the patient.
  • Use a well-fitting full-face mask (sized to the patient, small, medium or large).
  • Start at low pressures tolerated by most patients:
    • IPAP of 10 cm H2O
    • EPAP of 4 cm H2O
  • Other settings at default levels.
  • Allow the patient to hold the mask over their face without the straps tightened.
  • Tighten the straps to form a seal around the face, but not so tight as to squash the membranes of the mask.