Normoventilation with 100% oxygen for 10 minutes | Normoventilation with 100% oxygen for 10 minutes Hypocapnia is excluded (arterial blood PaCO2 target > 4,7 kPa). |
---|---|
Verify that the patient is normovolaemic, so that fluid loss caused by e.g. diabetes insipidus has been corrected. | |
Disconnect the patient completely from the ventilator for the time of testing. | A PEEP valve must be used during the apnoea test to prevent significant formation of atelectasis.
|
Ensure that, in the absense of exhalation, no pulmonary hyperinflation is caused, and that the gas flow has a route out. | |
If the patient is not a potential lung donor, the apnoea test may be carried out by administering oxygen through a suction catheter in the trachea. | |
The oxygen flow in the intubation tube is maintained at a level that ensures oxygenation. | |
Observation of the patient | The patient is observed for at least 10 minutes to detect spontaneous respiratory movements. |
The blood pressure and heart rate levels are maintained stable during the test in order to ensure tissue perfusion.
| |
Establishing the cessation of patient's own breathing | The patient's own breathing is established as having ceased, if spontaneous breathing movements do not occur and, in a blood gas analysis performed after the apnoea test, the pCO2in arterial blood is at least 8 kPa. |
Interrupting the test | If spontaneous breathing occurs during the test, the test must be interrupted and the patient must be reconnected to the ventilator. |
If oxygen saturation, measured with pulse oximetry, decreases below 90% despite increasing the oxygen flow, the test must be interrupted. | |
Repeating the test | The test may be repeated after a few hours, following good preoxygenation and lung recruitment.
|