Authors: Karen Crisfulla, RN, MSN, CCNS, CCRN
In this chapter, youll learn:
A Look at Pacemakers
A pacemaker is an artificial device that electrically stimulates the myocardium to depolarize, which begins a contraction.
Pacemakers are usually used for (1) bradycardia caused by sinus node dysfunction (SND), medications, or atrioventricular (AV) block and (2) cardiac dyssynchrony because of advanced heart failure. The device may be temporary or permanent, depending on the patients condition. Pacemakers are commonly necessary for the aging heart and following myocardial infarction or cardiac surgery.
and the Beat Goes on
Pacemakers work by generating an impulse from a power source and transmitting that impulse to the heart muscle. The impulse flows throughout the heart and causes the heart muscle to depolarize. Pacemakers usually consist of three components: the pulse generator, the pacing leads, and the electrodes at the tips of the pacing leads.
Making the Pacer Work
The pulse generator contains the pacemakers power source and circuitry. The lithium battery in a permanent or implanted pacemaker is its power source and lasts about 10 years. The circuitry of the pacemaker is a microchip that guides heart pacing.
A temporary pacemaker, which isnt implanted, is about the size of large cell phone or a telemetry box and is powered by alkaline batteries. These units also contain a microchip and are programmed by a touch pad or dials.
A Stimulus on the Move
An electrical stimulus from the pulse generator moves through wires (referred to as pacing leads or leads) to the electrode tips. The leads for a pacemaker designed to stimulate a single heart chamber are placed in either the atrium or the ventricle. For dual-chamber, or AV, pacing, the leads are placed in both chambers, usually on the right side of the heart.
One Pole or Two
The electrodes—one on a unipolar lead or two on a bipolar lead—send information about electrical impulses in the myocardium back to the pulse generator. The pulse generator senses the hearts electrical activity and responds according to how it has been programmed.
A unipolar lead system is more sensitive to the hearts intrinsic electrical activity than is a bipolar system. A bipolar system isnt as easily affected by electrical activity outside the heart and the generator (e.g., from skeletal muscle contraction or magnetic fields). (See A look at pacing leads.)
Working with Pacemakers
On an ECG, youll notice the spike from a unipolar system right away. (See Pacemaker spikes.) It occurs when the pacemaker sends an electrical impulse to the heart muscle. That impulse appears as a vertical line or spike. The spike with a bipolar system can sometimes be difficult to see.
Depending on the position of the electrode, the spike appears in different locations on the waveform.
Permanent and Temporary Pacemakers
Depending on the patients needs, a permanent or a temporary pacemaker can be used to maintain heart rhythm. Pacing lead placement varies according to the patients specific needs.
Permanent Pacemakers
A permanent pacemaker is used to treat chronic heart conditions such as SND, AV block, or dyssynchrony. SND causing symptomatic bradycardia can be because of advanced age or drug therapy. AV block is commonly because of cardiomyopathy, cardiac surgery, or septal ablation. Dyssynchrony between all the chambers and within the left ventricle is caused by advanced cardiomyopathy. Permanent pacemakers are surgically implanted, usually under local anesthesia. (See Placing a permanent pacemaker.)
Pocket Generator
The generator consists of a battery and microchip circuitry enclosed in a titanium case. It is implanted in a pocket made from subcutaneous tissue. The pocket is usually constructed under the clavicle. Permanent pacemakers are initially programmed during implantation. The programming determines how the pacemaker functions and can be adjusted externally when necessary.
Permanent Pacing Leads
Permanent pacing leads are usually placed transvenously. They are positioned in the appropriate chambers and then anchored to the endocardium or in the coronary sinus for left ventricular pacing. Some patients may require permanent epicardial pacing leads that are placed during cardiac surgery. This is common for patients with tricuspid valve surgery and for patients who require left ventricular pacing but a coronary sinus lead was unable to be placed transvenously. (See Placing a permanent pacemaker.)
Leadless Pacemakers
This is a self-contained pacemaker the size and shape of a large capsule of medicine. It is placed within the right ventricle via the femoral vein. Therefore, there is no chest incision or lead wire. It paces and senses only the right ventricle. It is indicated for people in whom atrial pacing is not needed. For example, for patients with:
Temporary Pacemakers
A temporary pacemaker is commonly inserted in an emergency. The patient may show signs of decreased cardiac output, such as hypotension or syncope. The temporary pacemaker supports the patient until the condition resolves.
A temporary pacemaker can also serve as a bridge until a permanent pacemaker is inserted, for example, immediately following cardiac surgery or while awaiting resolution of a bloodstream infection. Temporary pacemakers are used for patients with heart block, sinus bradycardia, or low cardiac output. Several types of temporary pacemakers are available, including transvenous, epicardial, and transcutaneous.
Temporary pacemakers should have the batteries, pacing thresholds, and connections between the pacing leads and generator routinely checked. The insertion sites should be kept sterile and be regularly monitored. Unit-based guidelines need to exist for this routine care.
Going the Transvenous Way
The transvenous approach is used when inserting a temporary pacemaker in nonsurgical environments. In this method, the pacing leads are commonly inserted through the subclavian or internal jugular vein. The transvenous pacemaker is probably the most common and reliable type of temporary pacemaker. Its usually inserted at the bedside or in a fluoroscopy suite. The pacing lead is typically advanced through a catheter into the right ventricle and then connected to an external pulse generator.
Taking the Epicardial Route
Epicardial pacemakers are commonly placed while patients are undergoing cardiac surgery. The doctor attaches the tips of the pacing leads (thin wires) to the surface of the heart and then brings the wires through the chest wall, below the incision. Theyre then attached to an external pulse generator. The pacing wires are usually cut or removed several days after surgery or when the patient no longer requires them.
Following the Transcutaneous Path
Use of an external or transcutaneous pacemaker has become commonplace. In this noninvasive method, one electrode pad is placed over the heart on the patients anterior chest wall, and a second is applied behind the heart on the back. An external pulse generator then emits pacing impulses that travel through the skin, muscle, and bone to the heart muscle. Transcutaneous pacing is also built into many external defibrillators for use in an emergency. In this case, the pacing electrode pads are built into the same electrode pads used for defibrillation.
Transcutaneous pacing is a quick and effective method of pacing heart rhythm and is commonly used in an emergency until a transvenous pacemaker can be inserted. However, some alert patients cant tolerate the painful electrical stimuli produced from pacing at the levels needed to capture the myocardium through the chest wall.
Setting the Controls
When your patient has a temporary pacemaker, youll notice several types of settings on the pulse generator. The rate control regulates how many impulses are generated in 1 minute and is measured in pulses per minute (ppm). The rate is usually set at 60 to 80 ppm. (See A look at a pulse generator.) The pacemaker is usually programmed in a demand mode. Thus, it fires only if the patients heart rate falls below the rate (e.g., 60) set on the pacemaker. The rate may be set higher in an effort to increase cardiac output or if the patient has a tachyarrhythmia thats being treated with overdrive pacing.
Measuring the Output
The electrical output of a temporary pacemaker is measured in milliamperes (mA). First, the stimulation threshold, or the minimum amount of energy required to capture or stimulate the cardiac muscle to depolarize, is assessed. Then the pacemakers output is set 10% higher than the stimulating threshold to ensure myocardial capture under various physiologic conditions, for example, hypokalemia.
Sensing the Norm
You can also program the pacemakers sensing or ability to sense electricity (e.g., P waves and QRS complexes). Sensing is measured and set in millivolts (mV). Most pacemakers are programmed in a demand mode; therefore, only pacing when the hearts own electrical activity is not sensed. Pacing is inhibited when the hearts own P waves and QRS complexes are sensed. This may be referred to as demand pacing. So many patients will at times have a paced rhythm, an intrinsic rhythm, or a rhythm with both paced and intrinsic beats at any given time.
Ages and stages |
Pacemakers in adult patients Adult patients are more apt to have AV block. Therefore, they usually get a dual-chamber pacemaker. This allows for synchronous pacing and contraction between the atria and ventricles. Cardiac output and exercise tolerance are increased because of increased ventricular filling. Older adults are more prone to permanent atrial fibrillation. Because atrial pacing is not possible during atrial fibrillation, a single-chamber ventricular pacemaker is typical for these patients when they have bradycardia. |
Pacemaker Codes
The capabilities of permanent pacemakers may be described by a generic five-letter coding system, although three letters are more commonly used. (See Pacemaker coding system.)
Introducing Letter 1
The first letter of the code identifies the heart chambers that can be paced when necessary. These are the options and the letters used to signify those options:
Learning About Letter 2
The second letter of the code signifies the heart chamber in which the pacemaker senses the intrinsic activity:
Looking at Letter 3
The third letter indicates the pacemakers response to the intrinsic electrical activity it senses in the atrium or ventricle:
Figuring Out Letter 4
The fourth letter of the code describes rate modulation, also known as rate responsiveness or rate adaptive pacing:
Finally, Letter 5
The final letter of the code is rarely used but specifies the location or absence of multisite pacing:
Pacemaker Modes
The mode of a pacemaker indicates its functions. Several different modes may be used during pacing. Here are three of the more commonly used modes and their three-letter abbreviations. (A three-letter code, rather than a five-letter code, is typically used to describe pacemaker function.) Pacemaker rates vary by age (see Pediatric pacemakers) and other patient characteristics.
AAI Mode
The AAI, or atrial demand , pacemaker is a single-chambered pacemaker that paces and senses the right atrium. When the pacemaker senses intrinsic atrial activity, it inhibits pacing and resets itself. If it does not sense intrinsic atrial activity, it paces. Only the atria are sensed and paced.
Not in Block
Because an AAI pacemaker requires a healthy and functioning AV node and resulting consistent ventricular conduction, it isnt used in AV block. An AAI pacemaker may be used in patients with sinus bradycardia, as after heart transplant, or with sick sinus syndrome as long as the HisPurkinje system isnt diseased.
VVI Mode
The VVI, or ventricular demand , pacemaker paces and senses the ventricles. (See AAI and VVI pacemakers.) When it senses intrinsic ventricular activity, it inhibits pacing. This single-chambered pacemaker benefits patients with complete heart block and those needing rare or less frequent pacing. Because it doesnt affect atrial activity, it may be used for patients with a temporary pacemaker (e.g., postcardiac surgery) who dont need an atrial kick. Atrial kick is the extra 15% to 30% of cardiac output that comes from atrial contraction. The VVI mode is also common for patients with a permanent pacemaker who have long-stand ing persistent atrial fibrillation.
Unsynchronized Activity
If the patient has spontaneous atrial activity, the VVI pacemaker wont synchronize the ventricular activity with it. So there will be no relationship between the P waves and paced QRS complexes, and tricuspid and mitral regurgitation may develop. Sedentary patients who dont need the atrial kick, or patients with atrial fibrillation, may receive this pacemaker.
DDD Mode
A DDD pacemaker is used with AV block. Annual incidence of AVB development in people with SND is about 15%. Therefore, some patients with SND get a dual-chamber pacemaker. (See DDD pacemaker rhythm strip.) However, because the pacemaker possesses so many capabilities, it may be hard to troubleshoot problems. Its advantages include its
Home, Home on the Rate Range
Unlike other pacemakers, the DDD pacemaker is set with a rate range, rather than a single critical rate. It senses atrial activity and ensures that the ventricles track or respond to each atrial stimulation up to an upper ventricular rate limit, thereby maintaining normal AV synchrony. A typical rate range is from 60 to 120 bpm. This causes pacing in the atrium once the atrial rate falls below 60. The ventricles will track (be paced after each P wave) if there is no sensed QRS complex, up to 120 bpm. A pacemaker cant stop a hearts intrinsic conduction. Therefore, even though a pacemakers upper rate may be 120, the patients heart rate may well exceed 120 bpm when there is intact intrinsic AV conduction.
Taking the Low Road
The DDD pacemaker paces (sometimes referred to as fires) in the atria when the atrial rate falls below the lower set rate. If the ventricles dont respond to an atrial event, the pacemaker will pace the ventricles. (See Evaluating a DDD pacemaker rhythm strip.) You or the telemetry monitor may calculate the heart rate by averaging 6 or 10 seconds. However, the pacemaker calculates the heart rate between two consecutive beats. Therefore, although you or a monitor may average the heart rate as being 60 or higher, the pacemaker will pace if the patients intrinsic rate is 59 between two consecutive beats.
Taking the High Road
The upper set rate acts as a safety mechanism by preventing the pacemaker from following (or tracking) an atrial tachyarrhythmia such as atrial fibrillation, atrial tachycardia, or atrial flutter. That upper rate limit is usually set at 120 or 130 bpm. Ones AV node would normally block the approximate 500 atrial bpm during atrial fibrillation from passing to the ventricle. The AV node is bypassed with a pacemaker, making the upper rate limit necessary with a DDD pacemaker.
Evaluating Pacemakers
Now youre ready to find out if your patients pacemaker is working correctly. To do this, follow the procedure described below.
Read the Records
First, determine the pacemakers mode and settings, particularly the rate setting. If your patient had a permanent pacemaker implanted before admission, ask them whether they have a wallet card from the manufacturer that notes which generator and pacing leads were implanted.
If the pacemaker was recently implanted, check the patients records for information. Dont check only the ECG tracing—you might misinterpret it if you dont know the mode or which pacing leads the patient has.
Look at the 12-Lead ECG
Next, review the patients 12-lead ECG. If it isnt available, examine lead V1 or MCL1 on telemetry. If there is only one ventricular lead, it is usually in the right ventricle. Therefore, expect a negatively deflected paced QRS complex in V1 or MCL1, just as with a left bundle branch block. An upright QRS complex in V1 or MCL1 may mean that the pacing lead is out of position, perhaps even perforating the septum and lodging in the left ventricle, or that there is a properly positioned pacing lead in the coronary sinus vein. When the His bundle is paced, the QRS complex should be narrow.
Scrutinize the Spikes
Then select a monitoring lead that clearly shows the pacemaker spikes. Some monitors have a filter or band width menu that allows adjustment for seeing larger pacemaker spikes. Make sure to use the paced mode on the telemetry monitor. If the monitor doesnt know its looking at a rhythm that may have pacemaker spikes, it can interpret a pacemaker spike as a QRS complex. Therefore, a pacemaker spike that is not followed by a QRS complex without the paced mode on will NOT cause an important alarm.
Mull over the Mode
When looking at the ECG tracing of a patient with a pacemaker, consider the pacemaker mode. Then interpret the paced rhythm. Does it match what you know about the pacemaker?
Unravel the Rhythm
Look for information that tells you which chamber is paced. Is there capture? Is there a P wave or QRS complex after each atrial or ventricular spike? Or do the P waves and QRS complexes stem from intrinsic activity?
Look for information about the pacemakers sensing ability. Do you see spikes where you wouldnt expect them? Like within or too close after a P wave or QRS complex? Remember if the pacemaker is sensing the electricity causing the P waves and QRS complexes, it should withhold a spike. Look at the rate. Whats the pacing rate per minute? Is it appropriate given the pacemaker upper and lower rate setting? Although you can determine the rate quickly by counting the number of complexes in a 6-second ECG strip, a more accurate method is to count the number of small boxes between two complexes and divide this into 1,500.
Troubleshooting Problems
Pacemaker malfunction can lead to arrhythmias and /or loss of AV synchrony. (See When a temporary pacemaker malfunctions) If you see questionable pacemaker activity for a patient with a permanent pacemaker, call the cardiology or cardiac electrophysiology team. There are actions you should take for patients with temporary pacemakers who have questionable pacemaker activity. Pacemaker problems that can lead to low cardiac output, hypotension, syncope, and death include
Mixed signals | |||||||
When a temporary pacemaker malfunctions Occasionally, pacemakers fail to function properly. When that happens, you need to take immediate action to correct the problem. The strips shown below are examples of problems that can occur with a temporary pacemaker and corrective actions to take in response.
|
Failure to Capture
Failure to capture is indicated on an ECG by a pacemaker spike without the appropriate atrial or ventricular response—a spike without a complex. Think of failure to capture as the pacemakers inability to stimulate the chamber to contract.
Causes include hypoxia, acidosis, an electrolyte imbalance, fibrosis, an incorrect lead position, a low pacing output setting (milliampere with temporary pacemakers and volts with permanent pacemakers), depletion of the battery, a broken or cracked pacing lead, or perforation of the pacing lead through the myocardium. It can lead to light-headedness, hypotension, or asystole.
Failure to Pace
Failure to pace is indicated by no pacemaker activity or spike at the set rate on an ECG. The problem is caused by battery or circuit failure, cracked or broken pacing leads, loose connections, oversensing, or the pacing output set too low. It can also lead to light-headedness, hypotension, or asystole.
Undersensing
Undersensing is demonstrated by a pacemaker spike when intrinsic cardiac activity is already present. Undersensing causes the pacemaker to fire even if intrinsic electricity is present. Think of it as help being given when none is needed. With undersensing, spikes occur on the ECG where they shouldnt. Although they may appear in any part of the cardiac cycle, the spikes are especially dangerous with a temporary pacemaker if they fall on the T wave. This can cause R-on-T, resulting in VT, ventricular fibrillation, or torsades de pointes. The output of a temporary pacemaker is much higher than that of a permanent pacemaker. Therefore, undersensing with a temporary pacemaker as opposed to a permanent pacemaker can lead to a fatal arrhythmia. With a permanent pacemaker, the unneeded pacing would most likely result in early battery depletion.
The problem can be caused by millivoltage set too high, electrolyte imbalances, disconnection or dislodgment of a lead, improper pacing lead placement, edema or fibrosis at the electrode tip, drug interactions, or a depleted or dead pacemaker battery.
Oversensing
If the pacemaker is too sensitive, it can misinterpret other electricity as the hearts electricity and therefore withhold pacing. This extraneous electricity can be from myopotentials from nearby muscle movement, electrical events in a different chamber, electricity from nongrounded items, or certain equipment that have a strong magnetic or electrical field. You should wear rubber gloves when hand ling exposed temporary pacing wires. This will prevent you from transmitting electrical current from static or ungrounded equipment to the patient. Likewise, temporary pacing wires should be covered with something nonconductive, like the finger of a rubber glove. Oversensing causes inhibition of pacing when the patient actually needs it. The heart rate, AV synchrony, and cardiac output wont be maintained. This can cause asystole, hypotension, light-headedness, and /or syncope.
How You Intervene
Make sure youre familiar with different types and modes of pacemakers and how they function. This will save you time and worry during an emergency. When caring for a patient with a pacemaker, follow these guidelines.
Checks and Balances
Temporary Pacemakers
All Pacemakers
On the Alert
What to Teach the Patient and Significant Other (s)
Temporary pacemakers:
Permanent pacemakers: In addition to explaining need, function, and expectations, give patients with permanent pacemakers the following:
A Look at Biventricular Pacemakers
Biventricular pacing is used in the treatment of some patients with class III and IV heart failure, with both systolic heart failure and intraventricular conduction delay. Also called cardiac resynchronization therapy, biventricular pacing reduces symptoms and improves the quality of life of patients with advanced heart failure. It also records heart failure parameters such as those indicating the amount of fluid in the chest.
Two Ventricles, Three Leads
Unlike other pacemakers, a biventricular pacemaker usually has three leads rather than one or two: one to pace the right atrium, one to pace the right ventricle, and one to pace the left ventricle. Both ventricles are commonly either paced at the same time or the left ventricle is paced milliseconds before the right ventricle. This causes them to contract about the same time, increasing cardiac output.
Different Ventricles, Different Timing
Under normal conditions, the right and left ventricles contract simultaneously to pump blood to the lungs and body, respectively. However, in heart failure, the damaged ventricles cant pump as forcefully and the amount of blood ejected with each contraction is reduced. If the ventricular conduction pathways are also damaged, electrical impulses reach the ventricles at different times, producing asynchronous contractions. This condition, called intraventricular conduction defect, further reduces the amount of blood that the heart pumps, worsening the patients symptoms.
An Important Tip
Unlike traditional lead placement, the pacing lead for the left ventricle is placed into a branch of a cardiac vein via the coronary sinus. This vein lies on the exterior of the left ventricle. The left ventricle is paced through the wall of the vein. Because this pacing lead isnt anchored in place, lead displacement may be more likely to occur. (See Biventricular lead placement.)
Improves Symptoms and Quality of Life
Biventricular pacing produces an improvement in the patients symptoms and activity tolerance. Moreover, biventricular pacing improves left ventricular remodeling and diastolic function and reduces sympathetic stimulation. As a result, in many patients, the progression of heart failure is slowed and quality of life is improved.
Sympathetic Response
To compensate for reduced cardiac output, the sympathetic nervous system releases neurohormones, such as aldosterone, norepinephrine, and vasopressin, to boost the amount of blood ejected with each contraction. The resultant tachycardia and vasoconstriction increase the hearts demand for oxygen, reduce diastolic filling time, promote sodium and water retention, and increase the pressure that the heart must pump against. The effect on the patient is a worsening of symptoms. Resynchronization therapy can reduce symptoms of heart failure, decrease hospitalizations and morbidity, and improve quality of life.
Whos a Cand idate?
Not all patients with heart failure benefit from biventricular pacing. Cand idates should have both systolic heart failure and intraventricular conduction delay along with these characteristics:
Many patients who meet criteria for biventricular pacing also meet criteria for a defibrillator. Therefore, patients having defibrillators with biventricular pacing (biventricular implantable cardioverter-defibrillator [ICD]) is very common.
Caring for the Patient
Provide the same basic care for the patient with a biventricular pacemaker that you would for a patient with a stand ard permanent pacemaker. Specific care includes these guidelines:
What to Teach the Patient
Provide the same basic teaching that you would for the patient receiving a permanent pacemaker. Additionally, when a patient gets a biventricular pacemaker, be sure to cover these points:
A Look at Implantable Cardioverter-Defibrillators
An ICD is an electronic device implanted in the body to provide continuous monitoring of the heart for bradycardia, VT, and ventricular fibrillation. The device then administers either override (antitachycardia) pacing beats or shocks to treat the dangerous arrhythmia. Implanted defibrillators are indicated for patients who have had a myocardial infarction and have an EF of 30% or less, or who have had nonsustained VT and an EF of 40% or less.
The placement and procedure for an ICD is usually similar to that of a permanent pacemaker and is usually inserted in an electrophysiology laboratory.
What It is
An ICD consists of a programmable pulse generator, a shocking lead, and one or more pacing leads. The pulse generator is a small battery-powered computer that monitors the hearts electrical signals and delivers electrical therapy when it identifies an abnormal rhythm. The leads are insulated wires that carry the hearts signal to the pulse generator and deliver the electrical energy from the pulse generator to the heart.
Storing and Retrieving Information
An ICD also stores information about the hearts activity before, during, and after an arrhythmia, along with tracking which treatment was delivered and the outcome of that treatment. Devices also store electrograms (electrical tracings similar to ECGs). With an interrogation programmer, a specially trained practitioner can retrieve information to evaluate ICD function, battery status, and to adjust ICD system settings. Biventricular ICDs, like biventricular pacemakers, store information that aid in heart failure assessment.
Automatic Response
Todays advanced devices can detect a wide range of arrhythmias and automatically respond with the appropriate therapy, such as bradycardia pacing (both single- and dual-chamber), biventricular pacing, antitachycardia pacing, cardioversion, and defibrillation. ICDs that provide therapy for atrial arrhythmias, such as atrial fibrillation, are also available. (See Types of ICD therapies.) All ICDs will temporarily pace postshock if necessary. Those with transvenous lead wires can also function as a typical pacemaker for chronic bradycardia needs and deliver antitachycardia pacing.
How Its Programmed
When caring for a patient with an ICD, its important to know how the device is programmed. This information is available through a status report that can be obtained and printed when the specially trained practitioner interrogates the device. This involves placing a small piece of equipment on the skin, over the implanted pulse generator, to retrieve information. The stored information helps to evaluate any arrhythmias that the device was programmed to store, any therapy delivered, and device function.
Program information includes
Arresting Developments
Follow these guidelines if your patient experiences an arrhythmia:
Be Alert
Provide the same basic care for the patient with an ICD that you would for a patient with a permanent pacemaker. Frequently, patients who are biventricularly paced also have an ICD. Specific care for patients with ICDs includes the following:
What to Teach the Patient
A Look at Radiofrequency Ablation
Radiofrequency ablation is an invasive procedure that may be used to treat arrhythmias in patients who havent responded to antiarrhythmic drugs or cardioversion or cant tolerate antiarrhythmic drugs. In this procedure, bursts of radiofrequency energy are delivered through a catheter to the heart tissue to destroy the focus of the arrhythmia or block its conduction pathway.
Whos a Cand idate?
Radiofrequency ablation is commonly used in treating patients with atrial tachycardia, atrial fibrillation, atrial flutter, AV nodal reentry tachycardia (AVNRT), inappropriate sinus tachycardia, bypass tracts like WolffParkinsonWhite (WPW) syndrome, VT, and frequent premature ventricular contractions.
Understand ing the Procedure
The patient first undergoes an electrophysiology study to identify and map the specific areas of the heart that cause or support the arrhythmia. The ablation catheters are inserted into a vein, usually the femoral vein, and advanced into the heart where short bursts of radiofrequency waves destroy small targeted areas of heart tissue. The destroyed tissue can no longer conduct electrical impulses. Other types of energy may also be used, such as microwave, sonar, or cryo (freezing).
Hitting the Target
Arrhythmias can be categorized by being either focal or reentrant. Examples of focal arrhythmias include atrial tachycardia, inappropriate sinus tachycardia, premature atrial contractions, and right- or left ventricular outflow tract VT. The arrhythmia originates from one spot or foci. Reentrant arrhythmias include VT (other than outflow tract), atrial fibrillation, AVNRT, and accessory pathways like WPW syndrome. Reentrant arrhythmias are supported by abnormal conduction that typically is in a continuous circular path or circuit. Radiofrequency ablation is used to either destroy the source of a focal tachycardia or block the abnormal pathway of the reentrant conduction circuit. (See Destroying the source or the circuit)
If a rapid arrhythmia originates above the AV node, the bundle of His in the AV node may be purposefully damaged with ablation to block impulses from reaching the ventricles. This may be done to control the ventricular rate for patients whose intolerable atrial rhythm cant be converted to sinus with chemical or electrical cardioversion or ablation. A permanent pacemaker would be necessary in tand em with this ablation. Also, anticoagulation therapy may be needed to reduce the risk of stroke.
If the patient has an accessory pathway as in WPW syndrome, electrophysiology studies are used to locate the accessory pathway and ablation can destroy the circuit.
After an ablation, ones heart rate and /or rhythm should return to normal. Thus, ablation can prevent tachycardia-mediated cardiomyopathy and may restore ventricular pumping function that has been damaged by the effects of prolonged tachycardia. Some ablations can be lifesaving, like VT ablations.
How You Intervene
When caring for a patient after radiofrequency ablation, follow these guidelines:
What to Teach the Patient
When a patient undergoes radiofrequency ablation, be sure to cover these points:
Subcutaneous implantable cardioverter-defibrillator
Test Strip
Time to try out a test strip. Ready? Go!
In the following ECG strip, the pacemaker is pacing and sensing the ventricles with 100% capture. The mode of response cant be evaluated because of lack of intrinsic activity. You would determine that the patient probably has:
a VVI pacemaker.
a DDD pacemaker.
an AAI pacemaker.
an AOO pacemaker.
Answer: B. The patient has a DDD pacemaker. It senses atrial activity and therefore inhibits atrial pacing. The P wave then triggers the pacemaker to try to sense a QRS complex. Because there is no intrinsic QRS complex after the P wave, the pacemaker paces the ventricle.
|
Selected References